Maintenance; Inspection Of The Sensor; Visual Inspection; Testing The Sensor With A Transmitter - Mettler Toledo InPro 5500i Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

InPro 5500 i CO
6

Maintenance

6.1

Inspection of the sensor

6.1.1 Visual inspection

To check your sensor, we recommend the following 
procedure:
–  The contacts of the connector must be dry. 
  M oisture, corrosion and dirt in the connector can 
lead to false readings.
–  Check the cable for buckling, brittle areas or 
  r uptures.
–  Before calibration always examine the Membra-
Cap optically for signs of damage. The Membra-
Cap must be intact and clean. Dirty membranes 
should be   w ip  e d clean using a soft, moist tissue.   
–  The membrane body has to be replaced once 
to twice per year to maintain max. measuring 
performance.
–  The MembraCap needs also to be replaced if the 
sensor   c annot be calibrated or the MembraCap 
shows sign of   m echanical damage.
–  The Intelligent Sensor Management (ISM)  
Dynamic Lifetime Indicator will tell the user when 
a MembraCap change needs to be performed.
a
Attention! Do not use any aggressive cleaning 
agents. This could damage the sensor or lead to 
faulty measurement. If the dirt can not be removed 
with a moist tissue, use gently non-aggressive 
solvent (alcohol). Any other chemicals shall not be 
used. This could damage the sensor or lead to a 
faulty measurement.

6.1.3 Testing the sensor with a transmitter 

For verifying the sensor, check measurements vs. 
reference (or known) sample values. 
This can be done in-line in process solution or using 
CO
 gas where the partial pressure is known. For 
2
off-line / workshop verification only use a calibration 
gas of a known value.
If the measured values differ from the desired values, 
a new process or a calibration with reference gas 
might be performed. 
h
Note: When the MembraCap is exchanged the 
sensor needs to be operated at least 15 min. before 
performing a calibration.
If the sensor does not perform as expected after 
replacing the MembraCap send the sensor to your 
local METTLER TOLEDO representative for inspection.
© 06 /14 Mettler-Toledo AG   
Printed in Switzerland 
 Sensor  
2
19
InPro 5500 i 
30 035 328

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Co2 sensor

Table des Matières