Technical Information
Programmable Analog Output Diagrams
The controller give the error "gauge error" if the gauge connected is
not supported. The pressure can be read by dispaly or by serial line
(win. 224) or by REMOTE I/O on Programmable Analog Out.
If the Programmable Analog Output is related to the pressure
read by an external gauge, the controller doesn't replicate
exactly the signal coming from the gauge; the controller converts
the gauge signal in the range from 0 to 10V.
The conversion formul is the following:
V
out
Where:
V
g
V
out
V
g_min
values can be drawn from the gauge technical specifications).
Figure 23
224/297
V
V
g
g
_
min
V
V
g
_
max
g
_
min
is the gauge output voltage;
is the Analog Output voltage;
V
is the gauge voltage measuring range (these
÷
g_max
Gauge Connector