Technical Information
Window Description
WIN # READ
WRITE
143
Always
Always
144
Always
Always
147
Always
Always
157
Always
Always
167
Always
Stop
200
Always
-
201
Always
-
202
Always
-
203
Always
-
204
Always
-
205
Always
-
206
Always
-
211
Always
-
216
Always
-
221
Always
-
282/302
TYPE
NAME
N
External Fan Config
L
External Fan Setting
N
Vent valve opening time
N
Gas Load Type
L
Stop speed reading
N
Bus current
N
3phase voltage
N
Output power
N
Driving frequency
N
Body temperature
N
Pump status
N
Error code
N
controller temperature
N
controller temperature
L
Actual Setpoint Status
TwisTorr 305-IC User Manual / 87-901-058-01
MEANING
external fan configuration:
0 = Always On
1 = Serial (generic output serially controlled)
External fan setting
Vent valve opening time (expressed in 0.2sec; 0=infinite)
0 = molar mass ≥ 30 g/mol (e.g., Ar)
1 = 4 g/mol < molar mass < 30 g/mol (e.g., N2)
2 = molar mass ≤ 4 g/mol (e.g., H2, He)
It reads the pump speed while in stop mode
It reads the actual value of the bus current supplied by
controller to electric motor
It reads the actual value of the pk 3phase voltage supplied
by controller to electric motor
It reads the actual value of power supplied by controller to
electric motor
This is the actual exciting frequency
It reads the Body temperature
Real time pump status:
0 = Stop
1 = Interlock
2 = Ramp
3 = Autotuning
4 = Braking
5 = Normal
6 = Fail
This location has a bit significant:
11 = Run Up Time
10 = Body HW Over Temp
9 = RFU/Not used
8 = Rotor Locked
7 = Too Hight Load
6 = Fault DRV8305
5 = Vdc (supply) Over Voltage
4 = Thermalpad Overtemp
3 = Vdc (supply) Under Voltage
2 = Controller Over Temp
1 = Pump Over Temp
0 = No connection to pump
It reads the MOSFET bridge temperature
It reads the controller PCB temperature
Reports the actual state of setpoint output