Calibration/Span Adjustment

Calibrate the Scale
Step
a. Switch on the scale
b. Zero the scale
c. Select calibration/linearization
mode
d. Select calibration
e. Confirm/start calibration.
After the zero point is stored,
the required calibration weight
is displayed
f. Place the required weight
on the scale
g. If the weight is applied within
the defined time limit and
tolerance, the "OK" symbol
is displayed.
Press "(Print/Enter)" to confirm
and save calibration.
(To exit calibration press
(Zero/Left) > 2 sec)
Key (or instruction)
(On/Off)
(Zero/Left)
(Tare/Right) > 2 sec
(Fn/Up) or (G/N/Down)
repeatedly
(Print/Enter)
(Print/Enter)
Display
21