Reassembling The Angled Screwdriver - KLS Martin group Angulus2 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 86
Instructions for Use
Angulus2 Angled Screwdriver
Ensure that the angled screwdriver is completely dry, both inside and out, following thermo
washer disinfection.
Remove any liquid residues using compressed air (maximum operating pressure of 3 bar).
Oil the dry angled screwdriver immediately after thermodisinfection.
6.2.5

Reassembling the Angled Screwdriver

Reassemble the angled screwdriver after manual and mechanical cleaning and disinfection.
NOTICE
Risk of material damage!
All 3 EN (serial numbers) must match. Possible damage to products in the case of a failure to
comply.
Insert the head into the sheath until it engages.
Pay attention to the symbols and turn from
Revision 04
to
until it locks.