Changing the bulb on the KaWe PICCOLIGHT®
E50: Unscrew the ophthalmoscope head from the
handle and pull the light source (5) out with your fin-
gers. If necessary, clean the glass bulb (7) of the new
lamp with alcohol. The glass bulb should be clean and
free of fingerprints and grease. Insert the new bulb
such that the side centring pin (6) rests in the groove in
the bulb socket. The bulb must be inserted completely
into the socket. Only connect the ophthalmoscope
head to the handle when the bulb is in the head. If
the bulb is not in place, a short circuit could occur.
KaWe PICCOLIGHT® E55
5 apertures
KaWe PICCOLIGHT® E55 / 2.5 V
with threaded screw-on connector
Direct ophthalmoscopy with 5 apertures for all oph-
thalmological examinations/19 correction lenses; from
-20 up to +20 dioptres/aspherical precision optics/
visual angle ±6°/aspheric precision optics for optimal
image reproduction quality/aspheric condenser for
optimal illumination.
Changing the bulb on the KaWe EUROLIGHT®
E55: Same as E50.
KaWe EUROLIGHT® E16 / E36 LED
KaWe PICCOLIGHT® E56 LED
6 apertures EU
KaWe EUROLIGHT® E36
KaWe PICCOLIGHT® E56
6 apertures US
KaWe EUROLIGHT® E16 / 2.5 V
with plug-in connector with fastening screw
Operating the KaWe EUROLIGHT® E16: Proper
operation of the instrument is guaranteed when origi-
nal KaWe battery/charging handles with 2, 1.5V, "C"
batteries or KaWe 2.5V rechargeable batteries are used.
Before using the instrument for the first time, charge
the batteries completely. To charge the batteries, use
the KaWe MedCharge® 4000 charging station. When
inserting the batteries, be sure to heed the directional
arrows on the rechargeable battery tag. Insert the in-
strument head onto the handle and secure it by tight-
ening the fastening screw on the side. Turn the instru-
ment on by pressing and turning the rotary switch to
the left. The brightness can be regulated easily with the
rheostat. Turn the instrument off by turning the switch
back to the right to its original position. Change the
batteries by removing the base cap.
Direct ophthalmoscopy with 6 apertures for all oph-
thalmological examinations/19 correction lenses; from
-20 up to +20 dioptres/aspherical precision optics/
visual angle ±6°/aspheric precision optics for optimal
image reproduction quality/aspheric condenser for
optimal illumination.
Changing the bulb on the KaWe EUROLIGHT®
E16: Same as E10.
KaWe EUROLIGHT® E36 / 2.5 V / 3.5 V
with snap-on connector
Changing the bulb on the KaWe EUROLIGHT®
E36: Same as E30.
15