Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The SwiSto3 is intended for use in the electromagnetic environment specified below. The customer or user of the SwiSto3 is to assure that it
is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5
GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location
in which the SwiSto3 device is used exceeds the applicable RF compliance level above, the SwiSto3 device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should be used no closer
to any part of the SwiSto3 device, including cables, than the recommended
separation distance, which is calculated using the equation for the frequency
transmission.
3 Vrms
Recommended separation distance
d = 1.2√P
3V/m
d = 1.2√P for 80 MHz to 800 MHz
d = 2.3√P for 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in meters (M).
Field strengths
netic site survey, should be less than the compliance level
range.
Interference may occur in the vicinity of equipment marked with the following
symbol:
25
from fixed RF transmitters, as determined by an electromag-
a
b
in each frequency