Technical Information
RS 232/RS 485 Communication Description
N.
Read/
Write
109
W
110
R/W
111
R/W
117
R/W
120
R/W
121
R/W
122
R/W
123
Reserved to Agilent service
124
172/182
Data
Description
Type
L
Pump life/ cycle time/ cycle number reset
L
Interlock type
(default = 1)
L
Analog output type: output voltage signal
proportional to frequency or power
N
Low Speed
frequency (Hz)
N
Rotational frequency setting (Hz)
N
Maximum rotational frequency in Hz (active
only in Stop condition)
L
Set vent valve on/off
(on = closed)
Controller TV70 Navigator User Manual / 87-900-956-01 (G)
Admitted
Values
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
250 to "Maximum
rotational
frequency"
(win 121)
(default = 833)
250 to "Maximum
rotational
frequency"
(win 121)
(default = 1250)
250 to 1250
(default = 1250)
On = 1
Off = 0
(default = 1)