The annealing tube must be clean and dry, inside and outside,
!
before it is installed in the sterilization orifice (4a).
The annealing tube is a consumable item. Check it regularly for
i
damage. Replace it if opacity, cracks, chips or similar are observed.
Quarz glass annealing tube:
Never operate the device without the annealing tube (4) in place!

6.2 Cleaning the glass front

The glass front can be cleaned with commercial glass cleaners and a soft, lint-free
cloth.
Never clean the glass front with abrasive sponges or abrasive cleaners be-
!
cause this may cause damage to the surface.

6.3 Cleaning the device

Proceed as in section 6. Additional remove power cord from the electrical
!
outlet.
It can be cleaned with a damp cloth and commercially available cleaning agents.
Care should be taken when cleaning that no liquid gets inside the unit. It is possible to
clean the appliance with a spray-and-wipe disinfectant. The instrument itself must not
be sterilized.
!
Flaming the device is not possible.
Depending on the residual heat, after switching off, the fan (8) runs until the
i
excess heat has been dissipated.

7. Fan

Smart fan control ensures that even in continuous operation the temperature of the
housing does not rise above 65 degrees. As an additional safety mechanism, there is
a mechanical thermal circuit breaker.
!
Do not cover the ventilation grilles (7) and cooling fan (8) on the unit.
Depending on the residual heat, after switching off, the fan (8) runs until the
i
excess heat has been dissipated.
Art.-No.: 4.000.400
EN
23