6. MAINTENANCE AND CARE

Tray cleaning
Maintaining tray cleanliness aids in maintaining correct functioning of the
device.
It is recommended to clean the internal part of the tray once a week using
a mild, chlorine-free detergent that does not react with aluminium. After
cleaning, the tray must be thoroughly washed with water.
Dry the tray before reinstalling the tray and push it over the front face pins
and push it down gently to lock it.
Cleaning the process chamber
Maintaining the chamber cleanliness aids in maintaining correct function-
ing of the device.
It is recommended to clean the process chamber interior once a week us-
47
MOUNTING STUDS
ing a mild, chlorine-free detergent. After cleaning, wipe the chamber with
a soft cloth until dry.
To clean the tray well it must be removed from the front of the device. To
do this, lift the tray gently up and pull it away from the front. The mounting
studs have notches in which the drawer fits.
Cleaning external surfaces
The external parts of the device should be cleaned using a soft cloth slight-
ly wet with water and a mild detergent (chlorine-free and not reacting with
plastics, varnishing coats and aluminium). Do not use strong detergents.
Use of mild detergents for maintaining the device does not affect the pos-
sibility of hazard related to toxic agents forming in contact with elements
of the device.
Cleaning the seal
It is recommended to clean the seal after 100 performed processes. Use
warm water and a microfibre cloth (microfibre with silver particles is ac-
ceptable) to clean the device. Use of dull or sharp cleaning tools is not
acceptable. Cleaning with chemical agents is not acceptable. Perform the
cleaning when the device has cooled down, after opening the drawer.
Maintain caution and do not bend the drawer. After cleaning, leave the
device open until the seal dries. During this time, protect the device from
damage. After cleaning and drying the seal, it can be lubricated with a sil-
icone lubricant.
Replacement of elements subject to wearing down
Elements subject to wearing down should be replaced periodically to en-
sure failure-free operation of the steriliser.