Preventing Device Damage - Seca 665 Mode D'emploi

Masquer les pouces Voir aussi pour 665:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 101
Preventing device
damage
Make sure that the patient does not step onto
and off the weighing platform right at the
edges.
Make sure that the patient steps onto and off
the weighing platform slowly and safely.
WARNING!
Risk of slipping
Ensure that the weighing platform is dry
before the patient steps onto it.
Ensure that the patients feet are dry before he
or she steps onto the weighing platform.
Make sure that the patient steps onto and off
the weighing platform slowly and safely.
NOTICE!
Damage to device
Ensure that no liquids enter the device. They
can damage the electronics.
Switch off the device before disconnecting
the power supply unit from the mains socket.
Disconnect the power supply unit from the
mains socket if you intend to not use the de-
vice for a longer period of time. Only this way
it can be ensured that the device is current-
less.
Make sure not to drop the device.
Do not expose the device to any impacts or
vibrations.
Perform function controls regularly as de-
scribed in the relevant section in this docu-
ment. Do not operate the device if it is
damaged or not working properly.
Ensure that there is no heat source in the im-
mediate vicinity. Do not expose to direct sun-
light. The excessive temperature could
damage the electronics.
Avoid rapid temperature fluctuations. When
the device is transported so that a tempera-
ture difference of more than 20 °C occurs, it
must stay turned off for at least 2 hours be-
fore it can be turned on again. Otherwise,
condensation water will form which can dam-
age the electronics.
English
Safety information • 59

Publicité

Table des Matières
loading

Table des Matières