Ambient conditions
During operation
During storage
Sampling
Respiratory volume
Breath flow
Duration of breath
Measurement accuracy
Standard deviation of the repeatability with ethyl alcohol standard
up to 0.50 mg/l
≥0.50 mg/l
up to 1.00 ‰
≥ 1.00 ‰
Sensitivity drift
Wait time for measurement readiness following a previous
measurement with:
0.25 mg/litre
0.50 mg/litre
1.00 mg/litre
>1.50 mg/litre 120 s
20 s
40 s
80 s
–5 to +50 °C
10 to 100 % rel. hum.
(non-condensing)
600 to 1300 hPa
–20 to +70 °C
>1.2 litres (preset)
>10 l/min (default)
>2 s (default)
0.008 mg/l
1.7 % of the measured value
0.017 ‰ or
1.7 % of the measured value
whichever value is higher
Typically 0.4 % of the measured
value per month
0.50 ‰
1.00 ‰
2.00 ‰
>3.00 ‰
Technical data
20 s
40 s
80 s
120 s