3.4
Procédures d'installation
Figure 3-2. Organigramme d'installation
START
HERE
Installation
Bench
Calibration?
BASIC SETUP
Set Sensor Type
Set Number of Wires
Set Units
Set Range Values
Set Damping
VERIFY
Simulate Sensor Input
Within
Specifications?
FIELD INSTALL
Set Failure Mode Switch
Mount Transmitter
Wire Transmitter
Power Transmitter
FINISHED
Installation
Septembre 2018