Monitor Relay Programming - Varian Turbo-V 301 Notice De Mode D'emploi

Table des Matières

Publicité

With the RS 232 connected, a bidirectional communi-
cation is established by selecting HOST. Data are sent
to an external computer every time the external
computer asks for the values.
The data available are:
Pump/controller operating condition
Cycle time
Pump life
Pump temperature
Pump current
Pump voltage
Controller output frequency
Cycle number
R1 condition
R2 condition
Life time and cycle # zeroing
Configuration parameter readings
Configuration parameter setting
If PRINT is selected and a printer is connected on RS
232 line, a unidirectional communication is established
and every minute the data are sent to the printer, even
if the pump is not running.
The set of data available are:
Pump speed KRPM
Pump temperature
Pump current A
Pump power W
R1 condition
R2 condition
The controller is factory-set to HOST.
Confirm the selection by pressing the PUMP
CURRENT push-button.
In this way you enter into an operating phase
named "Monitor Relay Programming" described in
the following paragraph.

Monitor Relay Programming

The display shows:
S P E E D
S E L E C T
where: XXKRPM = is the switch point of relay R1 at
the preset turbopump speed, adjustable from 00 to
99 KRPM.
The speed threshold will condition the R1 and R2
operation (see the following cycle diagram) and it is
factory-set to 53 KRPM.
Select the first number by pressing the CYCLE
NUMBER push-button, then enter the value by
pressing the PUMP CURRENT push-button.
Do
the
same
After pressing the PUMP CURRENT the second
time, the display will be as shown in the following
figure.
R U N - U P
S E L
:
X X h
where: RUN-UP TIME = is the interval time from
start to speed threshold value in hours, minutes,
seconds. Select from 00 to 99 hours, and from 00 to
59 minutes or seconds.
Select the run-up time according to the chamber
volume and/or operating cycle feature by pressing
the CYCLE NUMBER push-button to select the
desired number, then press the PUMP CURRENT
push-button to enter the data.
The run up time is factory-set to:
00h 08m 00s.
When the last digit is entered, the display will be as
shown in the following figure.
80
TECHNICAL INFORMATION
T H R E S H O L D
I O N : X X K R P M
for
the
second
T
I M E
X X m
X X s
87-900-940-01 (F)
number.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9437969-9537

Table des Matières