Maximum operating pressures—columns up to 9.4 mm id
Column Type
Poroshell 120
Poroshell 300
ZORBAX Rapid Resolution (RR)
ZORBAX Rapid Resolution High
Throughput (RRHT)
ZORBAX Rapid Resolution High
Definition (RRHD)
All ZORBAX, Pursuit, and Polaris
columns, not noted as RRHT or
RRHD, and HC-C18(2), TC-C18(2)
InfinityLab PFAS delay column
Mobile phase selection and operating
temperatures
The bonded stationary phase is nonpolar in nature and is
best used with polar mobile phases such as methanol/water
or acetonitrile/water mixtures. Increasing the amount of
organic component typically reduces the retention time of the
sample.
Particle Size
1.9 µm
2.7 μm, 4 μm
5 μm
3.5 µm
1.8 μm
1.8 μm
3.0 μm, 3.5 μm,
5 μm
English
Pressure Limit
1300 bar (19,500 psi)
600 bar (9000 psi)
400 bar (6000 psi)
400 bar (6000 psi)
600 bar (9000 psi)
1200 bar (17,000 psi)
400 bar (6000 psi)
200 bar (3000 psi)
(Metaguards)
1200 bar (17,000 psi)