• Fingers which are too thin or too cold may affect the measurement accuracy. Please clip
the thicker finger, such as thumb and middle finger deeply into the probe.
• The pulse oximeter can be used for adult and pediatric patients, based on the manufacturer
supplied finger probe.
• The update period of data is less than 5 seconds, and is subject to change based on varia-
tions in individual pulse rates.
• Please read the measurements on the waveform screen once the readings have stabilized.
this measure value is most accurate.
• If some abnormal conditions appear on the screen during test process, pull out the patient
finger and reinsert to restore normal use. if continued abnormal conditions appear, contact
your equipment provided.
• The device has a normal life expectancy of three years since the first electrified use.
• This device has alarm functions. Please reference chapter 6.1 for details.
• The device may not work for all patients. If you are unable to achieve stable readings, dis-
continue use.