recommended separation distances between portable and mobile rF
communications equipment and the eQuIPMent or systeM for eQuIPMent
or systeM that not LIFe-suPPOrtInG
portable and mobile rF communications equipment and the Pulse Oximeter (18720)
The Pulse Oximeter (18720) is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the Pulse Oximeter (18720) can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Pulse Oximeter (18720) as recommended below, according to the maximum output power of the communications equipment.
rated maximum output power of
transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
note 1 at 80mhz and 800mhz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
36
English
recommended separation distances between
Separation distance according to frequency of transmitter (m)
80mhz to 800mhz
0.1167
0.3689
1.1667
3.6893
11.6667
800mhz to 2.5ghz
0.2334
0.7378
2.3334
7.3786
23.3334