Table 4 Error list for pH and ORP sensors (continued)
Error
Description
TEMP
The temperature sensor is
MISSING
missing
GLASS IMP
The glass bulb is broken or
LOW
reached end of life
6.6 Warning list for sensors
A warning does not affect the operation of menus, relays and outputs. A
warning icon flashes and a message is shown on the bottom of the
measurement screen. To show the sensor warnings, press the MENU
key and select Sensor Diag, [Select Sensor], Warning List. A list of
possible warnings is shown in
Table
Table 5 Warning list for pH and ORP analog sensors
Warning
Description
PH TOO HIGH
The measured pH is > 13
ORP TOO HIGH
The measured ORP value
is > 2100 mV
PH TOO LOW
The measured pH is < 1
ORP TOO LOW
The measured ORP value
is < –2100 mV
OFFSET TOO
The offset is > 8 (pH) or
HIGH
200 mV (ORP)
OFFSET TOO
The offset is < 6 (pH)
LOW
or –200 mV (ORP)
20 English
Resolution
Examine the wiring for
the temperature sensor.
Make sure that the
correct temperature
element is selected.
Replace the sensor. Call
technical support.
5.
Resolution
Calibrate or replace
the sensor.
Calibrate or replace
the sensor.
Follow the
maintenance
procedures for the
sensor and then
repeat the
calibration.
Table 5 Warning list for pH and ORP analog sensors (continued)
Warning
Description
SLOPE TOO
The slope is >
HIGH
60 (pH)/1.3 (ORP)
SLOPE TOO
The slope is <
LOW
54 (pH)/0.7 (ORP)
TEMP TOO HIGH
The measured
temperature is >100 °C
TEMP TOO LOW
The measured
temperature is < 0 °C
CAL OVERDUE
The Cal Reminder time
has expired
REPLACE
The sensor has been in
SENSOR
operation > 365 days, or
the predicted life has
expired (Refer to
diagnostic and test menu
on page 18)
NOT
The sensor has not been
CALIBRATED
calibrated
FLASH FAILURE
External flash memory
failed
A ELEC TOO
The active electrode
HIGH
impedance is > 800 MΩ
A ELEC TOO
The active electrode
LOW
impedance is < 15 MΩ
Resolution
Repeat the
calibration with a
fresh buffer or
sample.
Clean the sensor,
then repeat the
calibration.
Make sure the right
temperature element
is used.
Calibrate the sensor.
Replace the sensor.
Sensor
Calibrate the sensor.
Contact technical
support.
The sensor is in air.
Return the sensor to
the process.
The sensor is
damaged or dirty.
Call technical
support.