3. Specifications
3.3 Voltage supply
The permissible power supply depends on the corresponding output signal.
Output signal
4 ... 20 mA
DC 0 ... 10 V
DC 0 ... 5 V
DC 1 ... 5 V
DC 0.5 ... 4.5 V ratiometric
Total current consumption: maximum 10 mA (except for 2-wire signals)
3.4 Accuracy
Accuracy
Accuracy at reference conditions
Non-linearity
Temperature error
Long-term drift (per IEC 61298-2)
Settling time
WIKA operating instructions, model MG-1
Power supply
DC 8 ... 30 V
DC 14 ... 30 V
DC 8 ... 30 V
DC 8 ... 30 V
DC 5 ± 0.5 V
≤ ±2 % of span
Including non-linearity, hysteresis, zero offset and end value deviation (corresponds to
measured error per IEC 61298-2). Calibrated in vertical mounting position with process
connection facing downwards.
≤ ±0.5 % of span BFSL (per IEC 61298-8)
≤ ±2.0 % of span
≤ 0.3 % of span/year
≤ 2 ms
EN