Calibration; Calibration Using The Remote Controller; Calibration Using Keys Of Crane Scale - Gram C3 Série Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CALIBRATION

Calibration using the remote controller

1. Power on the crane scale and keep pressed the D key until display shows
"CAL SP".
2. The display will show "CAL 00", press A key once to display "CAL00".
3. Do not hang on the hook any goods at this time.
4. Display will show "LOAD", press the A key until view "-------" and then
later on display "LOAD".
5. Press the D key to input the calibration weight value (long press D key to
increase the number of digit, click to the left shift).
6. Lifting weights, until steady 4 to 5 seconds.
7. Press the A key to show "--------", 2 seconds after the display of the
hanging weight values, calibration is finished, put weight down, turn it off.

Calibration using keys of crane scale

1. When power on and self detecting, keep pressed
shows "CAL SP".
2. Press the
3. Do not hang on the hook any goods at this moment.
4. Press the
the display will read "LOAD".
key once the display shows "CAL00".
key once the display shows "--------", after few seconds,
key until display
9

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

C3-300C3-500C3-1t

Table des Matières