STANDARD CALIBRATION

Press the
and
ECF-1
Weight calibration (Zero and Span)
ECF-2

Zero calibration

ECF-3

Span calibration

Weight calibration ECF-1
1. Press the
return to normal weighing mode.
2. The display will show CALZ, press the
3. Use the keys ◄, ►, ▲, 0 ~ 9 to introduce the calibration weight value and confirm
pressing the
4. Place the calibration weight on the platter and press the
5. The balance will automatically calibrate and return to normal weighing mode.
Zero calibration ECF-2
1. Press the
return to normal weighing mode.
2. The display will show CALZ, press the
3. The balance will automatically perform zero calibration and return to normal
weighing mode.
Span calibration ECF-3
1. Press the
return to normal weighing mode.
2. Use the keys ◄, ►, ▲, 0 ~ 9 to introduce the calibration weight value and confirm
pressing the
4. Place the calibration weight on the platter and press the
5. The balance will automatically perform span calibration and return to normal
weighing mode.
.
key at the same time to access to standard calibration.
key to access to mode ECF-1 or press the
key to access to mode ECF-2 or press the
key to access to mode ECF-3 or press the
key.
key.
key.
key.
key to exit and
key to exit and
key to exit and