OHAUS Frontier FC5515 Manuel D'instructions page 156

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

DE-32
FC5515
Rotor-Number
30130870; 24x1.5ml/2.0ml
30130871; 24x1.5ml/2.0ml
30130872; 30x1.5ml/2.0ml
30130879; 44x1.5ml/2.0ml
30130873; 12x5ml
30130874; 4x8place PCR stripes
30130881; 24 Hematocrit
12.6 Tabelle 6: Fehlermitteilungen
Error-No.:
Description
1
Imbalance arose
2
Imbalance sensor is defective
4
Imbalance switch has been activated for longer than 5 seconds
8
Transponder in the rotor is defective
11
Temperature sensor is defective
12
Chamber over temperature
14
Leap of speed is too big between two mesaurements
CLOSE lid
33
Open lid while motor is running
34
Lid contact defective
38
Lid motor is blocked
40
Communication with frequency converter distrubed during start
41
Communication with frequency converter distrubed during stop
42
Short circuit in the frequency converter
43
Undervoltage frequency converter
44
Overvoltage frequency converter
45
Over temperature frequency converter
46
Over temperature motor
47
Over current frequency converter
48
Timeout between control unit and frequency converter
49
Other error frequency converter
55
Overspeed
70
Timeout between controler and RS232 interface
99
Rotor is not allowed in this centrifuge
Inserted rotor does not exist in the programm
FALSE
Rotor is not detected
rotor no
Acceleration values
level 0
level 9
152
21
151
21
235
29
136
18
148
24
150
17
120
15
in seconds
Acceleration time
-1
from 0 min
-> U
max
TM
Frontier
Zentrifuge FC5515/FC5515R
Deceleration values
level 0
level 9
173
18
217
21
258
27
133
17
320
17
95
20
120
15
Deceleration time
-1
from U
-> 0 min
max