Balluff BMD 1L K Série Notice D'utilisation page 160

Sonde de mesure tige et câble avec qualification sil
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

7 Set up with the display and adjustment module
Diagnostics - Peak val-
ues, Additional
Diagnostics - Echo curve
46
If you have selected interface measurement under the menu item
"Setup - Application", the peak values of the interface measurement
are displayed in addition to the peak values of the level measurement.
In another window you can carry out a reset of the two peak values
separately.
The respective min. and max. measured values are saved in the
sensor. The values are displayed in the menu item "Peak values Ad-
ditional".
This menu item displays the peak values of the electronics tempera-
ture as well as the dielectric constant.
In another window you can carry out a reset of the two peak values
separately.
Information:
available.
The menu item "Echo curve" shows the signal strength of the echoes
over the measuring range in V. The signal strength enables an evalua-
tion of the quality of the measurement.
With the following functions you can zoom part sections of the echo
curve.
"X-Zoom": Zoom function for the meas. distance
"Unzoom": Reset the presentation to the nominal measuring range
V"

Publicité

Table des Matières
loading

Table des Matières