Recommended Disinfecting Solutions - GENDEX GXS-700 Manuel D'utilisateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 51
CAUTION
Autoclave sterilizers will permanently damage the GXS-700 sensor.

Recommended Disinfecting Solutions

Decontaminate the sensor, cable and holders in accordance with CDC – or your
country's (e.g., OSAP) standards – for infection control.
Gendex recommends wiping the GXS-700 sensor with one of the following cleaning
agents.
DO NOT:
The following solutions can be used to decontaminate the holders:
The GXS-700 sensor holders can be sterilized in a steam autoclave using distilled
water at temperature of 273
chemical disinfectants (refer to Instructions for Use). However, certain restrictions
apply:
Exposing the holders to hot metal and placing them close to heating elements will reduce the
lifetime of the holders.
032-0279-EN Rev 1
®
CaviCide
(Manufactured by Metrex)
CaViWipes
(Distributed by Kerr)
Asepticare (Manufactured by Ecolab)
®
Sani-Cloth
Plus (Distributed by Crosstex)
CIDEX OPA (Ortho-Phthalaldehyd, Distributed in the US by Advanced
Sterilization Products)
FD322 (Manufactured by Dürr)
Dürr System-Hygiene FD 350 Disinfection wipes Classic
(Manufactured by Dürr)
Isopropyl Alcohol
Clean the sensor using non appropriate tools.
o
o
F (134
C) at 216 kPa when bagged or using cold
Run the appropriate autoclave cycle according to the
manufacturer's specifications for your specific unit.
Always avoid direct contact of the holders with metal trays,
instruments, and heating elements.
Always place holders in autoclave bags.
Always situate bagged holders into the autoclave so that they are
furthest from the heat source.
Do not use a chemical autoclave.
Use
6-5