3. Selecting switch-off criterion

A switch-off criterion defines under which conditions the device should cancel the
drying process. Switch-off criteria save continuous time controls and manual
cancelling of drying processes. Furthermore they ensure that measurements are
always finished under the same conditions, thereby providing repeatable
measurements.
 After acknowledgement of the selected heating profile by
set heating profile and switch-off criterion flashes.
 Using arrow keys
by
.
Selectable settings suitable as switch-off criteria include:
<AUTO>
Automatic switch off
(2mg weight loss in 45 s)
<TIMING>
Time-controlled switch off,
selectable 3 min – 99 min
<MANUAL>
32
select required switch-off criterion and confirm
This switch-off criterion is based on a
weight loss per time unit. As soon as the
medium weight loss drops below the
desired value per unit of time, the
measurement will be finished
automatically.
When this switch-off criterion is selected
the measurement will continue until the set
drying time has lapsed
Manual switch-off by pressing stop key
, the currently
DAB-BA-e-1813