Begin temperature calibration:
Switch on temperature calibration set.
Close heating hood and press
heated to the set temperature. The current temperature and the elapsed time
will be indicated in the display.
After approx. 15 minutes finish by
the moisture analyser with that of the temperature calibration set. At a
difference of ±5°C we recommend to adjust the temperature, see following
chapter.
8.2.2
Temperature adjustment
If the admissible deviation is exceeded / not reached during temperature calibration
the temperature of the instrument can be adjusted as described below.
1. Switch off instrument.
2. Press
, quickly release, then press
„8.2X" will be displayed. If not, separate device from power supply and restart
step two.
3. Adjust temperature using arrow keys
button the temperature will be increased or reduced by 1°C.
4. Confirm input by
26
. The instrument will be automatically
. Compare the temperature value of
.
and
at the same time.
. Each time you press the
DAB-BA-e-1813