c. If the setup utility is not able to complete the firmware update,
contact technical support.

Troubleshooting

Problem
Possible cause
Communication
The communication
failure
wiring is not correct.
The instrument is not
configured.
There is an intermittent
connection problem.
1
Flow failure
A cap is on the sample
probe.
There are bends in the
tubing.
There is a tubing leak.
The vacuum pump has
failed.
22 English
Solution
Examine the system for loose
wiring/cables or incorrect
connections.
Configure the instrument. Refer
to
Configure the instrument
on page 14.
Use the CRTS (communication
reliable test software) to find
the problem if the FXB
communication protocol is
used.
Remove the cap from the
sample probe.
Examine the tubing for bends
that may cause an air flow
restriction.
Examine the tubing and fittings
for leaks.
Replace the vacuum pump
fuse if it is open. Refer to
Replace the fuse
on page 21.
Contact technical support.
Problem
Possible cause
2
Count alarm
There are high particle
counts in the room.
The probe is too near
the source of the
contamination.
The sensor may be
contaminated.
Sensor failure
The sensor is
contaminated.
1
A flow alarm temporarily turns the pump off for 30 seconds plus the length of
one sample. A flow alarm becomes inactive when the pump turns off.
2
A count alarm becomes inactive when the next sample starts.
Sensor decontamination
1. Use a portable particle counter to confirm the particle count
measured by the instrument.
If the particle count is similar, the problem is caused by the process
and not the instrument.
2. If the particle count is lower, clean the internal components. Refer to
Clean the interior surfaces
on page 13.
3. Do a purge to get a particle count of 0–1. Refer to
instrument
on page 13.
4. If the particle count does not go to 0–1, contact technical support.

Replacement parts and accessories

W A R N I N G
Personal injury hazard. Use of non-approved parts may cause
personal injury, damage to the instrument or equipment malfunction.
The replacement parts in this section are approved by the
manufacturer.
Solution
Troubleshoot the process to
identify the source of the
particles.
Change the position of the
probe.
Refer to
Sensor
decontamination
on page 22.
Purge the instrument. Refer to
Purge the instrument
on page 13.