User standard tables
1. Press SET.
2. Select [USER TABLES] and press OK.
3. Select [pH buffers] or [S/cm standards] and press OK.
4. Follow the instructions on the screen to enter the different values. The value
for 25°C is obligatory!
Channels
1. Press SET.
2. Select [CHANNELS] and press OK.
3. Select the desired ranges [INPUT] and press OK.
4. Select the desired inputs [ATC] (automatic temperature compensation) and
press OK.
5. Select [CONTROL] (yes or no) and press OK.
6. Select the control type (on/off or proportional) and press OK.
7. Pre-set the desired low level (Lo) and press OK.
8. Select the corresponding relay (REL) and press OK.
9. Pre-set the desired high level (Hi) and press OK.
10. Select the corresponding relay (REL) and press OK.
11. Pre-set the desired hysteresis (Hy) between the on and off state of the relays
and press OK.
12. Pre-set the desired delay (dt) after which the relays should be activated,
when one of the pre-set levels is exceeded and press OK.
Proportional control only
1. Pre-set the desired the proportional control area (dP), in which proportional
regulation should start and press OK.
2. Pre-set the desired pulse-time (Pt) of the relays and press OK. The relays will
pulse at a rate proportional to the regulation difference.
Optional alarm procedure
1. Select [ALARM] and press OK.
2. The display shows the alarm time after which the process control should be
interrupted in case of a failure. Pre-set the desired alarm-time and press OK.
3. Select the corresponding relay (REL) to be closed in case of an alarm and
press OK.
Optional maintenance procedure
1. Select [MAINTENANCE] and press OK.
2. The display shows the interval for a regular maintenance. Pre-set the desired
interval and press OK .
3. Pre-set the desired waiting period before the beginning of the maintenance
and press OK.
4. Select the corresponding relay (REL) and press OK.
5. Pre-set the maintenance period and press OK.
6. Pre-set the desired waiting period after the maintenance has finished and
press OK.
7. Pre-set the desired date and time when the first maintenance should start
and press OK.
The process control will automatically be interrupted during each maintenance
period!

Calibration reminder

Select the desired range by pressing é ê and then press CAL.
1. 1.
2. 2.
Select [GLP] and press OK.
3. 3.
Select [INTERVAL] and press OK.
4. 4.
Select the desired time interval between each automatic warning for
a new calibration of the electrodes and press OK.
3