1. Select the desired mode by pressing é ê . The display will immediately show
the measured value according to the previous calibration. Should you want to
recalibrate, press CAL.
2. The instrument permits to choose between the possible standards (1, 10, 100
ng/l, 1, 10, 100 µg/l, 1, 10, 100 mg/l, 1, 10, 100 g/l). Select the proper
values and press OK. The unused standards should be switched off.
3. Select [CALIBRATE] and press OK.
4. After rinsing the electrodes with distilled water, immerse them in the first
standard solution. Select the first calibration value and press OK when
readings are stable. Go on in the same way with the next standard solutions.
5. Decide whether a blank correction should be carried out or not. Follow the
instructions on the screen. After rinsing the electrodes with distilled water,
immerse them in a blank solution. When readings are stable press OK.
6. Rinse the electrodes with distilled water, immerse them in the samples, and
read the concentration on the display.
Maintenance: we refer to the manual supplied with the ion specific electrodes.
(R3620, R3630 only)
1. Press SET.
2. Select [DATA] and press OK.
3. Select [INTERVAL] and press OK.
4. Select the desired time interval between the data-logging and press OK.
View the stored values
1. Select [TABLE] and press OK.
2. Select the desired channel by pressing OK.
3. Select the first line to view and press OK.
4. Select [SHOW TABLE] and press OK.
Send the stored values
1. Select [SEND] and press OK to send the stored data to the digital port.
Erase the stored values
1. Select [ERASE] and press OK. Follow the instructions on the screen.
(R3620, R3630 only)
1. Press SET.
2. Select [RS485] and press OK.
3. Select [INTERVAL] and press OK.
4. Select the desired time interval between the data transmission and press OK.
5. Select [ID#] and press OK.
6. Select the desired identification number (0...999) and press OK. The
instrument can be identified by e.g. a computer when a specific number is
allocated to it.
7. Select [BAUD] and press OK.
8. Select the desired baud rate and press OK.
(R3620 only)

Data Storage in the internal memory

External storage in a computer using RS485