8 3 Handpiece / Cable

8 4 Cavitron 30K Ultrasonic Inserts

The many styles of Cavitron and Cavitron Bellissima 30K Ultrasonic Inserts are easily interchangeable for various procedures
and applications. See enclosed literature for specific information.
O-Ring
Provides seal for handpiece coolant.
O-ring should be replaced when worn.
Insert Tip
Shape and size of tip
determines access and
adaptation. Preheated
Lavage directed to tip.
Insert Marking
Finger Grip
Manufacturer,
Date (YYDDD=Double
digit year and triple digit
day of year), Frequency,
Type, Tip Lot Number
(if applicable).
Lavage Control
Turn the Lavage Control to select flow rate during system
operation. Flow rate is based on a scale from 1 to 6. Turn
clockwise toward 6 to increase flow at insert tip. Turn
counter-clockwise toward 1 to decrease flow. The flow rate
through the handpiece also determines the temperature of
the lavage. Lower flow rates produce warmer lavage. Higher
water flow rates produce cooler lavage.
If the handpiece becomes warm, increase the flow rate.
With experience the Dental Healthcare Professional will be
able to determine the best flow rate setting for optimum
operating efficiency and patient comfort.
Swivel Feature
Reduces cable drag as handpiece rotates during procedures.
Soft Nozzle Grip
Ergonomically designed to provide for a comfortable grasp
of the handpiece. The grip is a replaceable wear component.
Prior to use, verify that the grip is flush with the hard plastic
of the insert port.
Powder Delivery Port
Creates an airtight seal between the air polishing insert and
the handpiece. Replace when wear is noticed or powder is
leaking at nozzle interface.
Insert Port
The Cavitron
JET-Mate Sterilizable Handpiece accepts all
®
Cavitron
30K Ultrasonic Inserts and JET Air Polish inserts.
®
Connecting Body
Transfers and amplifies mechanical
motion of stack to insert tip.
Magnetostrictive Stack
Converts energy provided by the handpiece into
mechanical oscillations used to activate the insert
tip.