8.3 Handpiece / Cable

8.4 Cavitron JET Air Polishing Insert

Insert Heater Rod: Heats the
delivered water for patient
comfort.
Lavage Control
Turn the Lavage Control to select flow rate during system
operation. Flow rate is based on a scale from 1 to 6. Turn
clockwise toward 6 to increase flow at insert tip. Turn
counter-clockwise toward 1 to decrease flow. The flow rate
through the handpiece also determines the temperature of
the lavage. Lower flow rates produce warmer lavage. Higher
water flow rates produce cooler lavage.
If the handpiece becomes warm, increase the flow rate.
With experience the Dental Healthcare Professional will be
able to determine the best flow rate setting for optimum
operating efficiency and patient comfort.
Swivel Feature
Reduces cable drag as handpiece rotates during procedures.
Soft Nozzle Grip
Ergonomically designed to provide for a comfortable grasp
of the handpiece. The grip is a replaceable wear component.
Prior to use, verify that the grip is flush with the hard plastic
of the insert port.
Powder Delivery Port
Creates an airtight seal between the JET Air Polishing Insert
and the Jet-Mate™ handpiece. Replace when wear is noticed
or powder is leaking at nozzle interface.
Insert Port
The Cavitron
Polish inserts.
Prophy Powder Delivery Tube: Directs
air/powder flow to insert tip.
14
Jet-Mate™ Sterilizable Handpiece accepts JET Air
®
Insert Marking:
Manufacturer, Date (YDDD=Single digit
year and triple digit day of the year).
JET Air Polishing Insert Nozzle:
Tube-in-a-tube design delivers
precise air/water/powder mixture
at point of delivery.
O-Ring:
Seals water when Insert is fully seated in
Handpiece. O-ring should be replaced when
worn.