Start-Up Procedures; Between Patients; Shut-Down Procedures At The End Of The Day; 10 2 Weekly Maintenance - Cavitron Prophy-Jet Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
8. Turn the system ON.
9. If powder fluffing is observed when the Tap-On
Pedal is not in use, this would indicate an air leak. To
correct, turn the System OFF, remove the Powder Cap,
clean any residual powder from the O-ring seal and
threads,replace the Powder Cap, tighten, and turn the
System back ON.
10. Hold the sterilized Jet-Mate™ handpiece (without JET
Air Polishing Insert) over a sink or drain. Activate the
Purge Control button.
• The Purge button will light for two minutes indicating
proper activation of the purge function.
• If the Purge button is activated with a JET Air
Polishing Insert present in the Jet-Mate™ handpiece,
the button will blink for 3 seconds and disable.
Remove the JET Air Polishing Insert from the
Jet-Mate™ handpiece and press the Purge button
again.
• The Purge function can be interrupted at any time
during the two minute cycle by pressing the Purge
button again or by pressing the Tap-On
11. After completing the purge cycle, place a sterilized
JET Air Polishing Insert into the Jet-Mate™ handpiece
and adjust the Power Level Control to Prophy Mode,
and the Powder Flow and Lavage Controls to your
preferred operating positions.

BETWEEN PATIENTS:

1. Remove the used Cavitron
Clean and sterilize following the Infection Control
Procedures that were enclosed with your JET Air
Polishing Insert.
2. Hold the Jet-Mate™ handpiece over a sink or drain and
activate Purge function as described in Step 10 of the
Start-Up procedure.
3.
After the purge cycle is complete, turn the System
to the OFF (0) position.
4.
Remove the JET-Mate handpiece, clean and sterilize
following the procedures outlined in the Cavitron
Systems Infection Control Procedures booklet that was
enclosed with your system.
5. Disinfect the surfaces of the cabinet, Power Cord,
Handpiece Cable, Tap-On
TM
assembly (if applicable), Water Supply and Air Supply
lines by applying an approved non-immersion type
disinfectant solution* carefully following the instructions
provided by the disinfectant solution manufacturer.
To clean the system, generously spray disinfectant
solution on a clean towel and wipe all surfaces.
Discard used towel. Dry with a clean cloth. To disinfect
the system, generously spray disinfectant on a clean
towel and wipe all surfaces. Allow disinfectant solution
to air dry. Never spray disinfectant solution directly on
the system.
6.
Inspect the handpiece cable for any breaks or tears.
7.
If using a closed water supply or DualSelect Dispensing
TM
Foot
TM
Foot Pedal.
®
JET Air Polishing Insert.
Foot Pedal and Cable
18
System, check for adequate fluid volumes for the next
patient.
8.
Check the powder bowl for sufficient powder for the
next procedure.
9.
When ready to use, place a sterilized Jet-Mate™
Handpiece onto the handpiece cable assembly and
a sterilized JET Air Polishing Insert into the handpiece
and adjust system controls as preferred.
*NOTE: Broad spectrum hospital grade water-based
disinfection solutions are preferred, such as Lysol IC.
Some alcohol-based disinfectant solutions may be
harmful and may discolor plastic materials.
SHUT-DOWN PROCEDURES AT THE
END OF THE DAY:
1. Follow the "Between Patients" maintenance procedures,
steps 1 through 6. In addition, it is recommended to
close the manual shut-off valve on the dental water
supply system at the end of day.
2. Unscrew the powder bowl cap.
3. Remove the powder bowl from the unit and discard
the unused powder.
4. Holding the open end of the powder bowl away from
you, activate the system for 15 seconds to clear the
bowl. A high volume evacuator can be used to remove
any residual powder.
5. Remove the o-ring seal from the powder bowl cap and
using a soft dry cloth, wipe residual powder from the
cap, the o-ring and the powder bowl threads. Be
careful not to scratch or otherwise damage the cap.
6. Return the o-ring to the cap and secure the cap on the
powder bowl.
10.2 Weekly Maintenance
Remove residual prophy powder from the cap and bowl
threads using a soft brush (toothbrush). If not removed,
caked powder in threads can result in thread wear and
powder bowl cap disengagement.
It is strongly recommended that this system be
disinfected by chemically flushing the waterlines with
a 1:10 Sodium Hypochlorite solution (NaOCl) at the
end of each week. This can be accomplished by
connecting this device to the Cavitron DualSelect
Dispensing System or a number of other devices
available from your local distributors. Where this device
is connected to the Cavitron DualSelect Dispensing
System, please follow the DualSelect system's Directions
for Use manual. If connected to another device, please
follow those directions for use, keeping in mind that
a chemical flush should be performed at maximum water
flow for at least 30 seconds. The system should be left
undisturbed for 10 minutes but no more than 30 minutes

Publicité

Table des Matières
loading

Table des Matières