SECTION 8: Cavitron Plus Ultrasonic Scaler Description

8 1 System Controls

Ultrasonic Power Level Control
Turn knob to select the ultrasonic power level for operation. Turning the knob clockwise
increases the distance the insert tip moves (stroke) without changing the frequency;
turning the knob counterclockwise decreases the distance the insert tip moves (stroke)
without changing the frequency.
The Blue Zone is an extended low-power range for improved patient comfort when
subgingivally scaling.
Rinse
Turn the ultrasonic power level control knob fully counterclockwise until a "click" is heard.
Rinse mode is for use during an ultrasonic scaling procedure when lavage is desired with
minimal cavitation.
Blue Zone
Rinse
Tap-On
Wireless Foot
TM
Pedal
Eliminates the need to hold down
or pump foot pedal.
See section 8.6 for more details.
Main Power ON/OFF Switch
ON/OFF Switch located at the center front
underside of the system.
11
Diagnostic
Display
See section
8.2 for more
details.
Handpiece Holder
Securely holds the
system's handpiece,
or cable connector
when handpiece is not
installed.
Steri-Mate
®
Handpiece
Accepts all
Cavitron
30K
®
Ultrasonic inserts
See section 8.3 for
more details.