2.6
Ex-classification ATEX (not for InPro 6950i G)
2.6.1 Introduction
According to RL 94/9/EG (ATEX 95) Appendix l,
lnPro 6XXX*/*/*/*/* oxygen sensors are devices
group ll, category 1/2G and according to RL
99/92/EG (ATEX 137) may be used in zones 0/1 or
0/2 and gas groups llA, llB and llC that are potential-
ly explosive due to combustible substances in the
temperatures classes T3 to T6.
For use/installation, the requirements of EN 60079-
14 must be observed.
According to RL 94/9/EG (ATEX 95) Appendix l,
lnPro 6XXX*/*/*/*/* oxygen sensors are devices
group ll, category 1/2D and according to RL
99/92/EG (ATEX 137) may also be used in zones
20/21 that contain combustible dusts.
For use/installation, the requirements of EN 61241-
14 must be observed.
The digital sensor circuit is part of a common intrinsi-
cally safe system and is for operation connected to a
separately certified transmitter.
The digital sensor circuit as part of an intrinsically
safe system is isolated from the not-intrinsically safe
electric circuits up to a maximum rated voltage of
375 V and from grounded parts up to a maximum
rated voltage of 30 V.
2.6.2 Rated data
Measuring circuit:
Fail-safe ignition protection class Ex ia IIC only for
connection to a certified fail-safe circuit.
Maximum values:
U
i
L
= negligible
i
C
= negligible
i
Note:
The above maximum values are each the total of all
individual circuits of the associated intrinsically safe
power supply and transmitter.
InPro 6000 G – Series
52 206 477
InPro 6800 G / 6850i G / 6900i G / 6950i G O
16 V, I
30 mA, P
i
50 mW