Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

be used for the facilitation and rehabilitation of
motor faculties as well as a method to reduce
spasticity.
Intensity: depending on the muscle district to be
stimulated.
UROLOGY: Programmes recommended for the
stimulation of perineal muscles.
These programmes are recommended in case
of urological pain or incontinence.
Programmes to be used with probes in com-
pliance with the law.
NOTE: Because it includes clinical stimulation pro-
grammes, the apparatus is classified as a medi-
cal device. For this reason, it is certified accor-
ding to directive 93/42/EEC on medical devices
by CERMET, Notified Body No. 0476. The certifi-
cation encompasses clinical applications.
PROGRAMMABILITY (PRO models only)
To access programmability after turning on the
device, press the Fn key for 3 seconds. The first of
the 10 free programmes available appears. Use
the P+ or P- buttons to select the number of the
programme you wish to create and press OK to
confirm. By using the P+ and P- buttons, select
the type of impulse: continuous or intermittent
--- and press OK to confirm.
CONTINUOUS IMPULSE STIMULATION ---
a) Use the P+ or P- buttons to set the phase time
duration and press OK to confirm. (Pt)
b) Use the P+ or P- buttons to set the frequency
range and press OK to confirm. (Fw)
c) Use the P+ or P- buttons to set the work impul-
se width and press OK to confirm. (Aw)
To programme only phase 1, press the Fn key for
3 seconds. The programme will thus be saved.
Press Ok to begin the treatment. To programme
other phases, repeat stages a), b) and c). Once
all the phases have been programmed, press
the Fn key for 3 seconds.
INTERMITTENT IMPULSE STIMULATION
d) Use the P+ or P- buttons to set the phase time
duration and press OK to confirm. (Pt)
e) Use the P+ or P- buttons to set the frequency
range and press OK to confirm. (Fw)
f) Use the P+ or P- buttons to set the rest fre-
quency and press OK to confirm. (Fr)
g) Use the P+ or P- buttons to set the work/rest
impulse width and press OK to confirm. (Aw)
h) Use the P+ or P- buttons to set the work time
and press OK to confirm. (wt)
i) Use the P+ or P- buttons to set the rest time and
press OK to confirm. (rt)
To programme only phase 1, press the Fn key for
3 seconds. The programme will thus be saved.
Press Ok to begin the treatment. To programme
other phases, repeat stages d), e), f), g), h) and
i). Once all the phases have been programmed,
press the Fn key for 3 seconds.