METHOD OF USE

In order to connect the product to the user cor-
rectly, proceed as follows:
- connect the cables to the device cable soc-
kets;
- connect the electrodes to the relative cables
connected to the device;
- apply the electrodes to the skin.
The device and its accessories must be replaced
in the appropriate bag in compliance with the
following storage and transport conditions.
Temperature:
Air humidity:

DESCRIPTION OF THE DEVICE

Ch 1
F
A
C
E
A - Channel 1 intensity adjustment buttons
B - Channel 2 intensity adjustment buttons
C - Phase skip buttons (if present)
D - Buttons for programme number choice, time
modification and selection of parameters
E - ON/OFF - OK - PAUSE button
F - Function Key (Fn)
5-45 °C
20-85 %
Ch 2
B
D
Page 34 -
CAUTION !
For the ionophoresis treatment, use Channel 1 with the
specific electrodes only. Electrodes are available upon
payment.
Turn-on
To turn on the device, press the ON/OFF (OK)
button for three seconds until the acoustic signal
is heard. The software version/model is displayed
for three seconds with a number in the lower left
display.
Programme number choice.
Select the programme number using the P+ and
P- buttons. Press Ok to confirm.
During the programme number choice, the dis-
play shows the number of phases in the pro-
gramme (top) and the programme total dura-
tion (top right).
Start of the programme.
Once the programme has been selected,
increase channel intensity by using the channel
intensity adjustment buttons.
Display during execution.
During execution, the display shows the pro-
gramme number (top left), the total number of
phases and the phase in progress (full phase)
and the remaining time of the phase in progress
(top right).
Programme pause.
To pause the programme, press the OK button
on the Joypad. The time stops. Press OK again to
resume to the programme. Resumption after a
pause repeats the last interrupted cycle from the
beginning with set current values. During pause,
the blinking wording PAUSE appears on the dis-
play.
Programme interruption.
Should it be necessary to interrupt the program-
me, turn off the device by pressing the OK button
for three seconds.