Invacare Platinum Mobile Manuel D'utilisateur page 71

Concentrateur d'oxygène
Masquer les pouces Voir aussi pour Platinum Mobile:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

± 2 kV for power supply
Electrical Fast
lines
transient / burst
± 1 kV for input/output
IEC 61000-4-4
lines
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to ground
0% U
Voltage dips
and
IEC 61000-4-11
70% U
Single phase: at 0°
Voltage
Interruptions
0% U
IEC 61000-4-11
Power frequency
(50/60 Hz)
30 A/m
magnetic field
IEC 61000-4-8
1187376-C
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to ground
; 1 cycle
0% U
T
and
; 25/30 cycles
70% U
T
Single phase: at 0°
; 250/300 cycle
0% U
T
30 A/m
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
; 1 cycle
T
of the Device requires continued operation during
power mains interruptions, it is recommended that
the Device be powered from an un-interruptible
; 25/30 cycles
T
power supply or a battery.
U
is the a. c. mains voltage prior to application
T
of the test level.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Device requires continued operation during
power mains interruptions, it is recommended that
; 250/300 cycle
the Device be powered from an un-interruptible
T
power supply or a battery.
U
is the a. c. mains voltage prior to application
T
of the test level.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Technical Data
71

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Poc1-100c-c

Table des Matières