Invacare® Platinum® Mobile
4.5 Turning Off the Concentrator
1. To turn off the concentrator press the On/Off button
until "Powering Down" is shown on the display screen.
2. The power down sequence takes a approximately three
seconds.
4.6 Warm Up Period
CAUTION!
Risk of Minor Injury or Discomfort
During the warm up period (typically less
than five minutes) oxygen output is not within
specifications listed in Specifications in Technical
Data.
– Concentrator may be used during warm up
period.
After turning on the concentrator a warm up period of up to
15 minutes is needed to reach the specified oxygen purity.
"Warming Up" is shown on the display screen during this
period.
While warming up, the concentrator will automatically
deliver a pulse of oxygen about every four seconds when no
breath is detected. The nasal cannula should be connected
to the concentrator during warm up period.
When the concentrator has finished warming up, the current
pulse setting will be shown on the display screen. If no
breath is detected after warm up has ended, the No Breath
Detected alarm will occur. Refer to Alarm Conditions in
Troubleshooting.
36
4.7 Breathing with the Use of the Concentrator
As you breathe with the nasal cannula installed, a pulse
of oxygen is delivered each time the concentrator senses
an inhalation. Refer to Connecting/Positioning the Nasal
Cannula in Setup for cannula installation instructions.
If no breath is detected for 15 seconds, the No Breath
Detected alarm will occur. Refer to Alarm Conditions in
Troubleshooting.
4.8 Reading the Display Screen
Pulse Flow Setting and Pulse Icon
After the concentrator finishes warming up, the current
oxygen pulse flow setting B and pulse icon C are shown on
the display screen A.
The pulse icon flashes on each time an oxygen pulse is
triggered.
1187376-C