Resetting to default settings

You can reset the measurement parameters and the configuration to the de-
livery status separately from one another (initialization).
The following measurement parameters can be reset to the delivery status:
Conductivity (
InI)
Measuring mode
AutoRange, automatic switchover
of the measurement range
Cell constant
Temperature compensation
Temperature coefficient of the lin-
ear temperature compensation
TDS factor
The following configuration parameters (InI) can be reset to the delivery sta-
tus:
Baud rate
Interval 1
(automatically saved)
Interval 2
(for data transmission)
Reset procedure:
l To switch it off, press <CAL><t> while pressing the <RUN/ENTER> key.
l To reset the measurement parameters with < > < >, select YES and
confirm with <RUN/ENTER>, or
l Continue to the configuration parameters (InI) without resetting with the
<RUN/ENTER> key.
l After the configuration parameters, InI changes the instrument to the last
active measurement mode.
Resetting to default settings
On (YES)
-1
0.475 cm
(calibrated)
-1
0.475 cm
(manual range I)
-1
0.100 cm
(manual range II)
nLF
2.000 %/K
1.00
4800
OFF
OFF