Vercise™ Deep Brain Stimulation Systems
Troubleshooting
Telemetry Help
Common sources of interference, such as television and computer monitors, short range RFID
electronic tracking systems such as badge scanners and parking lot scanners, power adapters,
and wireless chargers, can affect wireless communication. You will continue to have access to the
Remote Control menus even if the Remote Control is outside its effective communication range
with the Stimulator. If you enter a command to change your settings or programs while the Remote
Control is outside of its effective communication range, you will have several seconds to move the
Remote Control within range of the Stimulator to send the command. After the command is received
by the Stimulator, the Remote Control will make a double-beep then display the requested change on
the screen. If the Remote Control is unable to communicate with the Stimulator, the Telemetry Error
message will appear:
To retry communication with the Stimulator – this resends the last command – select
Retry. To
cancel communication and return to the screen, press
.
Remote Control 3 Instructions for Use
92753326-04
40 of 203