Télécharger Imprimer la page

Boston Scientific Vercise Remote Control 3 Mode D'emploi page 38

Publicité

Vercise™ Deep Brain Stimulation Systems
Charge Stimulator Now or Disable MRI Mode Screen
If MRI Mode has already been enabled and the Stimulator battery power falls below the recommended
value, the Remote Control will display a message instructing the patient to charge the Stimulator.
To charge the Stimulator without disabling MRI Mode:
1. Do not press
.
2. Instruct the patient to charge the Stimulator.
3. Check the Remote Control to confirm that the error message has cleared.
4. Navigate to the Home screen on the Remote Control by pressing
Remote Control and confirm that the MR Conditional Symbol
screen.
The patient can also disable MRI Mode before charging the Stimulator:
1. Press
to disable MRI Mode.
2. Instruct the patient to fully charge the Stimulator.
3. Check the Remote Control to confirm that the error message has cleared.
4. Enable MRI Mode by following instructions in the "Enabling MRI Mode" section of this manual.
Warning:
The Charger and Remote Control are MR Unsafe and must not be brought into the
MRI scanner room.
Remote Control 3 Instructions for Use
92753326-04
32 of 203
on the side panel of the
is displayed on the Home

Publicité