Vercise™ Deep Brain Stimulation Systems
Yes to proceed with enabling MRI Mode or
3. Select
Warning:
An MRI scan is not recommended when the impedances are not within the acceptable
range. Higher or lower than normal impedances could indicate compromised
Stimulator‑Lead integrity. Scanning under these conditions may increase the risk
of potential adverse effects that are listed in the "Safety Information" section of
ImageReady
Stimulator Error Screen
If the system check fails due to a Stimulator error, MRI Mode will not be enabled and the Remote
Control will display the Stimulator Error screen. This screen may be displayed when there is a
communication error between the Remote Control and Stimulator. Press
message and retry entering MRI Mode. If this error screen immediately displays again, do not perform
an MRI scan and instruct the patient to contact the healthcare provider managing their DBS System or
Boston Scientific Technical Support.
Remote Control 3 Instructions for Use
92753326-04
34 of 203
MRI Guidelines for Boston Scientific DBS Systems.
TM
No to cancel the action.
to acknowledge the