Menu; Accessing The Menu; Parameter Settings - Tecno-gaz DAQF028 REV0.10 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TECNO-GAZ

8. MENU'

8.1 ACCESSING THE MENU

To access the menu keep the PRG key pressed for five seconds.
Press P01 and P02 to increase or decrease the programme index.
Press START to display a parameter setting or to return to the index of parameters.
Press P01 and P02 to increase or decrease the various parameters.
Press STOP to exit the parameters page.

8.2 PARAMETER SETTINGS

To set parameters will be required a password.
PARAM.NR
PARAMETER DESCRIPTION V1.5
P01
Set-up disinfection temperature cycle 2 and cycle 3
P02
Set-up disinfection temperature cycle 1
P03
Disinfection time cycle 3
P04
Max temperature heating element during disinfection
P05
Time 1° washing
P06
Time 2° washing
P07
Disinfection time cycle 2
P08
Loading time of limescale-remover
P09
Regeneration time ( water softner built-in)
P10
Drain time before disinfection
P11
Drain time during cooling
P12
Water loading time during cooling
P13
Detergent loading time
P14
Water loading time during regeneration (water softner built-
in)
P15
Maximum waiting time in lock door
P16
Lime-scale value
10 : no regeneration
20 : regeneration every 25 cycles
30 : regeneration every 18 cycles
40 : regeneration every 12 cycles
50 : regeneration every 6 cycles
60 : regeneration each cycle
P17
Drain time during last cooling period
P18
Maximum waiting time cold water loading
P19
Maximum waiting time of drain
P20
Maximum time of heating elements functioning
P21
Maximum waiting drain time during water loading
P22
Disinfection temperature fluctuation
P23
Loading time of detergent for filling up pipe
P24
Loading time of lime-scale remover for filling up pipe
P25
Do not modify this parameter . To be changed by
manufacturer only
P26
0= N°1 probe , 1= N°2 probes
P27
Hot washing temperature
P28
Disinfection time cycle 1
P29
Rising period short cycle (0=OFF ; 1=ON)
P30
Rising period standard cycle (0=OFF ; 1=ON)
P31
Acid detergent dosing pump(0) anti lime-scale dosing pump(1)
P32
Loading time of acid detergent
P33
Number of regeneration to load salt
UNIT
°C
°C
Sec.
°C
Sec.
Sec.
Sec.
Sec.
Sec.
Sec.
Sec.
Sec.
Sec.
Sec.
Sec.
FR°
Sec.
Sec.
Sec.
Min.
Sec.
°C
sec.
sec.
30
DEFAULT
FACTORY
SET UP
90
90
600
60
30
240
180
10
600
30
15
15
25
120
2
60
30
120
15
30
50
2
60
60
1
1
65
60
1
1
1
25
8
DAQF028_rev0.10
RANGE
85-93
55-93
300-600
1-180
30-120
30-600
180-300
0-30
480-900
0-60
0-60
0-60
0-60
40-150
0-10
10-60
10-30
0-240
0-60
0-60
0-60
0-10
0-60
0-60
0-1
0-1
40-80
1-180
0-1
0-1
0-1
0-60
1-18

Publicité