4.5.4 Measuring the Density of a Liquid Using a Calibrated Sinker

Press and hold the Function / Mode button until Density
appears. Press Yes to enter the Density Determination.
Item Settings:
Sample Type: Liquid
Volume: The calibrated sinker has a default volume of 10.0 ml,
which can be increased or decreased by pressing No or Back.
After setting the volume, press Yes to begin the weighing.
Note: when the Density Type is set to Liquid, the Liquid type
and Porous material selections are disabled.
1. Weigh the calibrated sinker in the air with the balance and
the density determination kit.
When the * (symbol of stability) appears, press the
Function / Mode button to confirm the weight of the
calibrated sinker in the air.