Nitrate T / M260
Implementation of the provision Nitrate with Tablet and Powder
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
Fill 24 mm vial with 10 ml
sample.
Zero
Press the ZERO button.
For devices that require no ZERO measurement , start here.
Fill a Nitratest tube with
20 ml sample.
21
10 ml
Close vial(s).
Remove the vial from the
sample chamber.
Add one microspoon NI-
TRATE TEST powder .
Sample
Place sample vial in the
sample chamber. • Pay at-
tention to the positioning.
Empty vial.
Close the test tube with the
lid and mix the contents
by vigorously shaking for
1 minute.
EN   Method Reference Book   1.0