Unpacking after centrifugation
6.8
Centrifugation
6.8.1
Centrifugation in continuous operation
6.8.2
Centrifugation with time preselection
AB5005en
5.
Place the connections over the support plate so that the valves
cannot break.
Ensure that the hoses do not project out of the insert.
Stow hose sections projecting over the edge of the insert between the
folded-down support plate and the insert wall.
6.
If necessary, balance weights should be placed between the folded-
down support plate and the receptacle wall.
1.
Pull the satellite bag out of the insert while holding the silicone plate in
place with one hand.
2.
Slowly pull out the folded-down part of the support plate at the loop
provided for this purpose.
Return the support plate to its original shape in a controlled manner.
The folded-down part of the support plate may spring back and mix
blood components.
3.
Remove the remaining blood bag either together with the support
plate or after removing the support plate from the insert.
Personnel:
■
Trained user
[t] button repeatedly until the input field of the 't/min:'
1.
Press the
parameter is reverse highlighted.
2.
Select the value 0 using the rotary knob.
[t] button repeatedly until the input field of the 't/ :sec'
3.
Press the
parameter is reverse highlighted.
4.
Select the value 0 using the rotary knob.
The input field displays
[START] button.
5.
Press the
The centrifugation run is started.
The
'Rotation' indicator is lit up while the rotor is turning.
The timing starts at 00:00.
The rotor speed or the resulting RCF value, the temperature in the
centrifuging chamber and the elapsed time are displayed during
the centrifugation run.
[STOP] button to cancel the centrifugation run.
6.
Press the
Ramp-down takes place with the set ramp-down parameters.
'OPEN OEFFNEN' is displayed.
Personnel:
■
Trained user
[t] button repeatedly until the input field of the 't/min:'
1.
Press the
parameter is reverse highlighted.
[Rotary knob] to set the desired value.
2.
Use the
Rev.: 20 / 01.2024
'---:--' .
Operation
31 / 50