4.1
Signal Integrity of the Transferred Signal
The following sections describe the effect that bandwidth and the type of connec-
tion have on signal integrity.
4.1.1
Bandwidth
The bandwidth BW of a probe is one of its specific parameters. The bandwidth of
the probe and the bandwidth of the base unit together form the system band-
width. The following explanations refer to the probe itself, but can also be applied
to the entire system.
Fig. 4-2: Amplitude frequency response of the R&S RT-ZZ80
The bandwidth
● Specifies the maximum frequency at which a purely sinusoidal signal is still
transferred at 70 % (–3 dB) of its amplitude.
See
figure 4-2
User Manual 1409.7689.02 ─ 02
Measurement Principles
Signal Integrity of the Transferred Signal
23