ThermoFisher Scientific BIOShield 1000A Mode D'emploi page 3

Table des Matières

Publicité

Centre of Emergency Preparedness and Response
Health Protection Agency
Porton Down
Salisbury
Wiltshire SP4 OJG
United Kingdom
Certificate of Containment Testing
Containment Testing of Thermo
Scientific rotor 75003603
Report No. 59-08 G
Report prepared for: Thermo Fisher
th
Issue Date: 15
January 2009
Test Summary
A Thermo Scientific 75003603 contained rotor (Max speed 6,030
rpm) was supplied by Thermo Fisher and containment tested at 6,030
rpm using the method described i n Annex M of EN 61010-2-020.
The rotor was shown to contain a spill when tested i n triplicate.
Report Written By
Report Authorised By

Publicité

Table des Matières
loading

Table des Matières