6. How to use
Press the ON/OFF switch.
By pressing once, exposure is started for 10 seconds.
The exposure time can be extended to 30 seconds by depressing the ON/OFF button repeatedly several times.
Press 2x = 20 seconds / press 3x = 30 seconds.
Hold the light conductor as close as possible to the region to be cured (4-7 mm). The light conductor must no contact the dental material
for the first few seconds. Once the first layer is sufficiently cured, the light conductor may come into contact with the dental composite.
The device is equipped with overheating protection, which is activated if the temperature reaches 50°C. When overheating protection is
activated, the colour indicator ring will flash yellow for the next 60 sec. and is then deactivated until the temperature of the handpiece
reaches 35°C.
For safety reasons never charge the device unattended. Therefore, please remove the plug when leaving the dental practice.
VI. Care and maintenance
1. Remove hardened plastic residues from light conductor. Do not use metallic, pointed or sharp objects to this purpose.
2. Disinfect the handpiece with a soft cloth and disinfectant solution. Please do not spray, only wipe. Liquids should not penetrate into the
interior of the handpiece. Do not use abrasive cleaning agents. We recommend M+W alcohol-free disinfection spray. Moisten the cloth and
carefully wipe the handpiece.