70
Laser class
Wavelength of therapy beam
Mode
Pulse length
Pulse mode
1
(single pulse)
2
(double pulse)
3
(triple pulse)
Energy attenuation
Beam diameter at the focus
Angle of exit aperture (divergence)
Aiming beam
NOHD
Rated voltage
Rated frequency
Rated current
Fuses
Electrical Protection Class
Protection type
Instrument type
Earth conductor
Control unit dimensions
Control unit weight

Technical Specifications

VISULAS YAG III laser system

IV (CFR 21, Part 1040, Sec. 1040.10)
4 (DIN EN 60825-1:2003)
1064 nm
Supergaussian
typ. 4 ns
Energy (typical)
Max. pulse repetition rate
10 mJ
2.5 Hz (5 pulse / 2 s)
23 mJ
1 Hz (1 pulse / s)
35 mJ
0.5 Hz (1 pulse / 2 s)
22 levels: 2, 4, 6, 8, 10, 12, 14, 16, 20, 24, 28, 32, 36, 40, 42, 48, 56, 60, 64,
70, 80 and 100 % transmission
2
10 µm in air (1 / e
)
16° (round angle)
Wave length: 660 to 680 nm
Power output: max. 150 µW
4-point aiming beam system for focusing
2 m
100 V to 240 V ± 10 %
50 to 60 Hz
Max 1.4 A to 0.7 A; transient: max. 2.8 A (< 1 s)
2x T 3.15 A / E, 5 x 20 mm in acc. with IEC 60127
SK I
IP 20
B (in acc. with IEC 60601-1)
The instrument should only be connected to properly earthed power outlets.
H x W x D = 135 mm x 210 mm x 330 mm
4 kg
Burst frequency
-
33 kHz
33 kHz